| Home > Technical
Notes > ICP-MS
Trace
Element Analyses of Rocks and Minerals by ICP-MS
Knaack,
C., Cornelius, S.B., and Hooper, P.R.,
GeoAnalytical Lab, Washington
State University, December, 1994
INTRODUCTION
The ICP-MS (inductively coupled plasma
source mass spectrometer) consists of a quadrupole mass spectrometer
with an inductively coupled argon plasma as an ion source. Liquids
introduced into the plasma (7000°C) are ionized and then passed
to the mass spectrometer through a two-stage ion extraction
interface. The ICP-MS is capable of quantitatively determining
trace elements in liquids in the range of fractions of a
part per billion. For routine REE analysis of rocks and minerals,
the detection limit is at or below chondrite levels. Its
capability for rapid multi-element analysis at low cost,
high sensitivity, and relative freedom from interferences
make the ICP-MS an excellent instrument for the determination
of many trace elements in rocks and minerals.
In the routine
procedure practiced in the GeoAnalytical Laboratory for trace
elements in rocks and minerals, the following 26 elements are analyzed:
all 14 naturally occurring rare earth elements (La through Lu)
together with Ba, Rb, Y, Nb, Cs, Hf, Ta, Pb, Th, U, Sr and Zr.
Zr is measured only as a check for complete dissolution of the
sample.
SAMPLE PREPARATION
Rock and mineral samples must
be completely dissolved prior to analysis. The digestion technique
used for routine analysis is adapted from Crock and Lichte (1982)
and our own experimentation. Samples are first ground in an iron
bowl in a shatterbox swing mill. Two grams of this rock or mineral
powder is then mixed with an equal amount of lithium tetraborate
(Li2B4O7) flux, placed into a carbon crucible and fused in a 1000° muffle
furnace for 30 minutes. The resulting fusion bead is briefly ground
again in the shatterbox and 250 mg of this powder is dissolved
on a hotplate at 110°C, using 6 ml HF, 2 ml HNO3, and 2 ml
HClO4 in an open teflon vial. The sample is evaporated to dryness,
followed by an additional evaporation with 2 ml HClO4 at 165°C
to convert insoluble fluorides to soluble perchlorates. 3 ml HNO3,
8 drops H2O2, 5 drops of HF and an internal standard of In, Re,
and Ru are added to the sample which is then is diluted up to 60
ml final volume (1:240 final dilution). This combined fusion/dissolution
procedure ensures the complete dissolution of zircons and other
refractory phases such as garnets, while removing silica and boron
as matrix elements by volitilizing them as gaseous fluorides.
Back to top
INSTRUMENT OPERATION
The instrumentation consists
of a Sciex Elan model 250 ICP-MS equipped with a Babington nebulizer,
water cooled spray chamber, and Brooks mass flow controllers. Samples
are introduced into the argon plasma at 1.0 ml/min using a peristaltic
pump and an automatic sampler. Plasma power is 1500 watts. Under
these conditions MO+/M+ (the proportion of metal ions forming oxides)
is minimized. The instrument is run in "multi-element" mode
averaging 10 repeats of 0.5 sec/element for a total integrated
count time of 5 sec/element.
Most elements have more than one isotope.
For these elements the selection of the isotope for measurement
is based on relative abundance and freedom from oxide and isobaric
interferences. Table 1 lists the isotopes measured and their
potential interfering oxide.
CALIBRATION
Unknown samples are run in sets of 17.
One acid blank and two samples each of the 3 in-house rock standards
BCR-P, GMP-01, and MON-01 (table 2) are run with each batch, totaling
24 standard and unknown samples per batch.
The three in-house standards
have been calibrated against 17 international standards (figure
1). Their elemental concentrations so derived are listed in table
2. Independent values on these three in-house standards by Walsh
(Kings College, London, U.K., by ICP-OES), Meier (U.S.G.S., Denver,
by ICP-MS), Bailey and Conrey (WSU, Pullman, WA, by INAA) are also
listed.
CALCULATION OF CONCENTRATIONS
Intensities for standards
and unknowns are downloaded to a personal computer and reduced
using a conventional spreadsheet program. Raw intensities are corrected
for oxide and isobaric interferences as listed in table 1 (modified
from Lichte et. al., 1987) and corrected for drift using the In,
Re, and Ru internal standards (after Doherty, 1989).
Calibration curves for each element are then constructed from the six (2x3)
standard samples and single acid blank by plotting given values (table 2)
against the corrected intensities. Concentrations for the unknown samples are
then computed from this curve. These calculations assume that the isotopic proportions
of the unknowns and standards do not vary significantly from the
average of the earth's crust.
Back to top
PRECISION
The reproducibility (total of sample preparation
plus instrumental precision) is illustrated in table 3, which
includes 24 separate preparations of BCR-P analyzed in 12 separate
runs between September 1994 and December 1994.
ACCURACY
As always, no absolute values for the standard
samples are available to determine the calibration curves; only
a consensus of the most likely values. It follows that exact measurements
of accuracy cannot be made. Perhaps the best way of estimating
analytical accuracy in secondary techniques, as employed here,
is to calculate the scatter of individual standards from a calibration
line which represents the best fit for all the standards.
This is
demonstrated visually for each element in the attached series
of 24 plots (Figure 1). Values for each of these standards, treated
as unknowns, have been calculated from these calibration curves
and are listed in table 2. This table also reproduces the recommended
values (Govindaraju, 1994) used in constructing the calibration
curves.
REFERENCES CITED
Crock, J.G., Lichte, F.E., 1982,
Determination of rare earth elements in geologic materials by inductively
couple argon plasma/atomic emission spectrometry. Analytical Chemistry,
54, 1329-1332.
Doherty, W., 1989, An internal standardization
procedure for the determination of yttrium and the rare earth elements
in geological materials by inductively coupled plasma mass spectrometry.
Spectrochemica Acta, 44B, 263-280.
Lichte, F.E., Meier, A.L., Crock,
J.G., 1987, Determination of rare earth elements in geological
materials by inductively coupled mass spectrometry. Analytical
Chemistry, 59, 1150-1157.
REFERENCES FOR TABLE 4
Govindaraju, K., 1994 compilation
of working values and sample descriptions for 383 geostandards.
Geostandards Newsletter, Special Issue, July 1994, 18, 15-53.
TABLES
AND FIGURES
Table 1: List of isotopes used in analyses.
Table 2: Trace element concentrations
of the 3 in-house standards used to develop calibration curves.
Table
3: Precision of a single sample (BCR-P) measured over a four
month period (September 1994 - December 1994).
Table 4: Estimate of accuracy
for ICP/MS analyses. WSU and recommended values for the international
rock standards used to calibrate the three in-house standards.
Figure
1: Corrected intensity versus recommended value for the international
standards used to calibrate the three in-house standards for
the 26 elements analyzed.
Back to top
For figures and more information, please contact
Charles Knaack |